组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Environmental estrogen pollution and estrogen effects on the female reproductive system are well recognized scientifically. Among the estrogens, 17 β-estradiol is a priority in environmental estrogen pollution, and it is also a major contributor to estrogen which regulates the female reproductive system. 17 β-estradiol is carcinogenic and has a tumor promotion effect relating to breast cancer, lung cancer and others. It also affects psychological well-being such as depression, fatigue and others. Thus, a simple method of detecting 17 β-estradiol will be important for both environmental estrogen pollution and health care. This study demonstrates a single-use, cost-effective 17 β-estradiol biosensor system which can be used for both environmental and health care applications. The bio-recognition mechanism is based on the influence of the redox couple, K₃Fe(CN)₆/K₄Fe(CN)₆ by the interaction between 17 β-estradiol antigen and its α-receptor (ER-α; α-estrogen antibody). The transduction mechanism is an electrochemical analytical technique, differential pulse voltammetry (DPV). The levels of 17 β-estradiol antigen studied were between 2.25 pg/mL and 2250 pg/mL; Phosphate buffered saline (PBS), tap water from the Cleveland regional water district, and simulated urine were used as the test media covering the potential application areas for 17 β-estradiol detection. An interference study by testosterone, which has a similar chemical structure and molecular weight as those of 17 β-estradiol, was carried out, and this 17 β-estradiol biosensor showed excellent specificity without any interference by similar chemicals.