2017

Design and Validation of a Biosensor Implantation Capsule Robot.

Journal of biomechanical engineering Xie W, Lewis WM, Kaser J, Ross Welch C, Li P, Nelson CA, Kothari V, Terry BS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We have proposed a long-term, noninvasive, nonrestrictive method of delivering and implanting a biosensor within the body via a swallowable implantation capsule robot (ICR). The design and preliminary validation of the ICR's primary subsystem-the sensor deployment system-is discussed and evidence is provided for major design choices. The purpose of the sensor deployment system is to adhere a small biosensor to the mucosa of the intestine long-term, and the modality was inspired by tapeworms and other organisms that employ a strategy of mechanical adhesion to soft tissue via the combined use of hooks or needles and suckers. Testing was performed to refine the design of the suction and needle attachment as well as the sensor ejection features of the ICR. An experiment was conducted in which needle sharpness, needle length, and vacuum volume were varied, and no statistically significant difference was observed. Finally, preliminary testing, coupled with prior work within a live porcine model, provided evidence that this is a promising approach for implanting a biosensor within the small intestine.

关键词