组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Herein, an electrochemical urea sensing bio-electrode is reported that has been constructed by firstly electropolymerizing 4-(2,5-Di(thiophen-2-yl)-1H-pyrrol-1-yl)aniline monomer (SNS-Aniline) on Pencil Graphite Electrode (PGE), then modifying the polymer coated electrode surface with di-amino-Ferrocene (DAFc) as the mediator, and lastly Urease enzyme through glutaraldehyde crosslinking. The effect of pH, temperature, polymer thickness, and applied potential on the electrode current response was investigated besides performing storage and operational stability experiments with the interference studies. The resulting urea biosensor's amperometric response was linear in the range of 0.1-8.5mM with the sensitivity of 0.54μA/mM, detection limit of 12μM, and short response time of 2s. The designed bio-electrode was tested with real human blood and urine samples where it showed excellent analytical performance with insignificant interference.