组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
A new type of carbon nanotube (CNT)-based impedimetric biosensing method has been developed for rapid and selective detection of live bacterial cells. A proof-of-concept study was conducted using T2 bacteriophage-based biosensors for electrochemical detection of Escherichia coli B. The T2 bacteriophage (virus) served as the biorecognition element, which was immobilized on polyethylenimine (PEI)-functionalized carbon nanotube transducer on glassy carbon electrode. Charge-directed, orientated immobilization of bacteriophage particles on carbon nanotubes was achieved through covalent linkage of phage capsid onto the carbon nanotubes. The presence of the immobilized phage on carbon nanotube-modified electrode was confirmed by fluorescence microscopy. Electrochemical impedance spectroscopy (EIS) was used to monitor the changes in the interfacial impedance due to the binding of E. coli B to T2 phage on the CNT-modified electrode. The detection was highly selective toward the B strain of E. coli as no signal was observed for the nonhost K strain of E. coli. The present achievable detection limit of the biosensor is 103 CFU/mL.