2017

Development of a progesterone immunosensor based on thionine-graphene oxide composites platforms: Improvement by biotin-streptavidin-amplified system.

Talanta Dong XX, Yuan LP, Liu YX, Wu MF, Liu B, Sun YM, Shen YD, Xu ZL
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Progesterone (P4) is a kind of hormone that can cause neuropathic disturbances in humans when the concentration overpasses a certain degree. In this work, an electrochemical immunosensor capable of detecting P4 sensitively and selectively was developed. Thionine-graphene oxide (Thi-GO) composites with excellent biocompatibility were synthesized and coated to a clear glassy carbon electrode. P4 coating antigen (P4-OVA) was immobilized to the electrode, then sample as well as biotinylated antibody (biotin-P4 Ab) were added. The free P4 can compete with P4-OVA for binding to biotin-P4 Ab. After the further addition of streptavidin-HRP, H2O2 was introduced to develop electrical signal for quantitative determination of P4. After careful optimization of assay conditions, the proposed immunosensor showed a linear range from 0.02 to 20ngmL-1 for P4 in milk samples. The averaged recoveries from spiked samples ranged from 84.0% to 102.0%, which correlated well with standard HPLC-MS/MS. The biosensor also showed good specificity, reproducibility and stability, indicating its potential application in monitoring of P4 in a simple and low cost manner.

关键词