组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Progesterone (P4) is a kind of hormone that can cause neuropathic disturbances in humans when the concentration overpasses a certain degree. In this work, an electrochemical immunosensor capable of detecting P4 sensitively and selectively was developed. Thionine-graphene oxide (Thi-GO) composites with excellent biocompatibility were synthesized and coated to a clear glassy carbon electrode. P4 coating antigen (P4-OVA) was immobilized to the electrode, then sample as well as biotinylated antibody (biotin-P4 Ab) were added. The free P4 can compete with P4-OVA for binding to biotin-P4 Ab. After the further addition of streptavidin-HRP, H2O2 was introduced to develop electrical signal for quantitative determination of P4. After careful optimization of assay conditions, the proposed immunosensor showed a linear range from 0.02 to 20ngmL-1 for P4 in milk samples. The averaged recoveries from spiked samples ranged from 84.0% to 102.0%, which correlated well with standard HPLC-MS/MS. The biosensor also showed good specificity, reproducibility and stability, indicating its potential application in monitoring of P4 in a simple and low cost manner.