Advanced healthcare materials Park S, Choi J, Jeun M, Kim Y, Yuk SS, Kim SK, Song CS, Lee S, Lee KH
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Current methods to detect avian influenza viruses (AIV) are time consuming and lo inw sensitivity, necessitating a faster and more sensitive sensor for on-site epidemic detection in poultry farms and urban population centers. This study reports a field effect transistor (FET) based AIV sensor that detects nucleoproteins (NP) within 30 minutes, down to an LOD of 103 EID50 mL-1 from a live animal cloacal swab. Previously reported FET sensors for AIV detection have not targeted NPs, an internal protein shared across multiple strains, due to the difficulty of field-effect sensing in a highly ionic lysis buffer. The AIV sensor overcomes the sensitivity limit with an FET-based platform enhanced with a disposable well gate (DWG) that is readily replaceable after each measurement. In a single procedure, the virus-containing sample is immersed in a lysis buffer mixture to expose NPs to the DWG surface. In comparison with commercial AIV rapid kits, the AIV sensor is proved to be highly sensitive, fast, and compact, proving its potential effectiveness as a portable biosensor.