2017

An amperometric H

Bioscience reports Narwal V, Yadav N, Thakur M, Pundir CS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The nanoparticles (NPs) of hemoglobin (Hb) were prepared by desolvation method and characterized by transmission electron microscopy (TEM), UV spectroscopy and Fourier-transform IR (FTIR) spectroscopy. An amperometric H2O2 biosensor was constructed by immobilizing HbNPs covalently on to a polycrystalline Au electrode (AuE). HbNPs/AuE were characterized by scanning electron microscopy (SEM), cyclic voltammetry (CV) and electrochemical impedance spectra (EIS) before and after immobilization of HbNPs. The HbNPs/AuE showed optimum response within 2.5 s at pH 6.5 in 0.1 M sodium phosphate buffer (PB) containing 100 μM H2O2 at 30°C, when operated at -0.2 V against Ag/AgCl. The HbNPs/AuE exhibited Vmax of 5.161 ± 0.1 μA cm-2 with apparent Michaelis-Menten constant (Km) of 0.1 ± 0.01 mM. The biosensor showed lower detection limit (1.0 μM), high sensitivity (129 ± 0.25 μA cm-2 mM-1) and wider linear range (1.0-1200 μM) for H2O2 as compared with earlier biosensors. The analytical recoveries of added H2O2 in serum (0.5 and 1.0 μM) were 97.77 and 98.01% respectively, within and between batch coefficients of variation (CV) were 3.16 and 3.36% respectively. There was a good correlation between sera H2O2 values obtained by standard enzymic colorimetric method and the present biosensor (correlation coefficient, R2 =0.99). The biosensor measured H2O2 level in sera of apparently healthy subjects and persons suffering from diabetes type II. The HbNPs/AuE lost 10% of its initial activity after 90 days of regular use, when stored dry at 4°C.

关键词