Continuous and simultaneous determination of venous blood metabolites.

Talanta Andersson A, Chen Q, Groop L, Bülow L, Xie B
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Metabolic syndrome is associated with cardiovascular disease, type 2 diabetes mellitus (T2DM) and prediabetes. Metabolic syndrome is a cluster of interrelated clinical disorders. Difficulties in regulating glucose levels in blood are implicated in many of these disorders. Lactate, another energy metabolite, is produced under anaerobic conditions and can be used to monitor the balance between aerobic and anaerobic metabolism. Tested together, these metabolite levels can provide pro-diagnostic information that improves patient outcomes. Glucose and lactate were determined continuously and simultaneously in whole blood using a dual-channel thermal biosensor device in which one channel employed glucose oxidase for glucose analysis in comparison with lactate oxidase for lactate analysis in the others. No detectable clogging or interference was observed using venous blood samples. The linear detection range for both the glucose and lactate assays was 0.5-45mM. The sampling rate of up to 24 samples per hour with assay cycle time of 2.5min was achieved. Comparative analysis between our device and the HemoCue method showed an excellent correlation. The device was stable for hundreds of injections over a period of 45 days. The broad linear range, fast response and detection sensitivity are satisfactory for the clinical requirements, e.g. for diabetic or cardiovascular patients in intensive care units or surgical operation, where the tight control of blood glucose can decrease morbidity or mortality.