组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Uranium is one of the most common radioactive contaminants in the environment. As a major nuclear material in production, environmental samples (like soil and groundwater) can provide signatures on uranium production activity inside the facility. Thus, developing a new and portable analytical technology for uranium in aqueous media is significant not only for environmental monitoring, but also for nonproliferation. In this work, a label-free method for the detection of uranyl (UO22+) ions is developed by monitoring the translocation of a peptide probe in a nanopore. Based on the difference in the number of peptide events in the absence and presence of uranyl ions, nanomolar concentration of UO22+ ions could be detected in minutes. The method is highly selective; micromolar concentrations of Cd2+, Cu2+, Zn2+, Ni2+, Pb2+, Hg2+, Th4+, Mg2+, and Ca2+ would not interfere with the detection of UO22+ ions. In addition, simulated water samples were successfully analyzed.