A highly sensitive immunosensor based on ITO thin films covered by a new semi-conductive conjugated polymer for the determination of TNFα in human saliva and serum samples.

Biosensors & bioelectronics Aydın EB, Aydın M, Sezgintürk MK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A novel, ultrasensitive impedimetric immunosensor was constructed for the detection of tumor necrosis factor α (TNFα) by using Poly(3-thiophene acetic acid) (P3), a conjugated polymer as an immobilization matrix. The polymer P3 contains a lot of carboxylic acid groups on its surface that provide a larger biorecognition surface. This developed immunosensor was prepared on hydroxy-bearing ITO surface via an ester bond linkage of polymer P3 to immobilize anti-TNF α antibodies. The ITO electrode modification steps and interaction between anti-TNF α antibody and TNF α antigen were monitored by cyclic voltammetry (CV) and by electrochemical impedance spectroscopy (EIS) method. After the analytical parameters optimization, a linear detection response from 0.01pg/mL to 2pg/mL, a limit of detection LOD of 3.7 fg/mL and a limit of quantification (LOQ) of 12.4 fg/mL were achieved, which provided accurate results (relative standard deviation; 4.03%). The characterization of this developed immunosensor was performed by using Fourier transform infrared spectroscopy (FTIR), scanning electron microscopy (SEM), SEM-energy dispersive X-ray (EDX) mapping and atomic force microscopy (AFM). The immunosensor allowed a simple and fast detection of TNF α antigen in human serum and satisfied recoveries (98.69-105.20%) were obtained by using standard addition method.