2017

Direct Electron Transfer-type Bioelectrocatalysis of Peroxidase at Mesoporous Carbon Electrodes and Its Application for Glucose Determination Based on Bienzyme System.

Analytical sciences : the international journal of the Japan Society for Analytical Chemistry Xia HQ, Kitazumi Y, Shirai O, Kano K
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Non-catalytic direct electron transfer (DET) signal of Compound I of horseradish peroxidase (POD) was first detected at 0.7 V on POD/carbon nanotube mixture-modified electrodes. Excellent performance of DET-type bioelectrocatalysis was achieved with POD immobilized with glutaraldehyde on Ketjen Black (KB)-modified electrodes for H2O2 reduction with an onset potential of 0.65 V (vs. Ag | AgCl | sat. KCl) without any electrode surface modification. The POD-immobilized KB electrode was found to be suitable for detecting H2O2 with a low detection limit (0.1 μM at S/N = 3) at -0.1 V. By co-immobilizing glucose oxidase (GOD) and POD on the KB-modified electrode, a bienzyme electrode was constructed to couple the oxidase reaction of GOD with the DET-type bioelectrocatalytic reduction of H2O2 by POD. The amperometric detection of glucose was performed with a high sensitivity (0.33 ± 0.01 μA cm-2 μM-1) and a low detection limit (2 μM at S/N = 3).

关键词