Study of Alzheimer's Disease-Related Biophysical Kinetics with a Microslit-Embedded Cantilever Sensor in a Liquid Environment.

Sensors (Basel, Switzerland) Chae MS, Kim J, Yoo YK, Lee JH, Kim TG, Hwang KS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A microsized slit-embedded cantilever sensor (slit cantilever) was fabricated and evaluated as a biosensing platform in a liquid environment. In order to minimize the degradation caused by viscous damping, a 300 × 100 µm² (length × width) sized cantilever was released by a 5 µm gap-surrounding and vibrated by an internal piezoelectric-driven self-actuator. Owing to the structure, when the single side of the slit cantilever was exposed to liquid a significant quality factor (Q = 35) could be achieved. To assess the sensing performance, the slit cantilever was exploited to study the biophysical kinetics related to Aβ peptide. First, the quantification of Aβ peptide with a concentration of 10 pg/mL to 1 μg/mL was performed. The resonant responses exhibited a dynamic range from 100 pg/mL to 100 ng/mL (-56.5 to -774 ΔHz) and a dissociation constant (KD) of binding affinity was calculated as 1.75 nM. Finally, the Aβ self-aggregation associated with AD pathogenesis was monitored by adding monomeric Aβ peptides. As the concentration of added analyte increased from 100 ng/mL to 10 µg/mL, both the frequency shift values (-813 to -1804 ΔHz) and associate time constant increased. These results showed the excellent sensing performance of the slit cantilever overcoming a major drawback in liquid environments to become a promising diagnostic tool candidate.