2017

A colorimetric aptamer biosensor based on cationic polythiophene derivative as peroxidase mimetics for the ultrasensitive detection of thrombin.

Talanta Liu M, Li J, Li B
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A colorimetric assay for the ultrasensitive determination of thrombin was presented, in which the cationic polythiophene derivative was used as catalyst of the 3,3',5,5'-tetramethylbenzidine (TMB)-H2O2 reaction and the thrombin-binding aptamer (TBA) was used as inducing polymer's different conformation elements. It was found the cationic polythiophene derivative, poly[3-(3'-N,N,N-triethylamino-1'-propyloxy)-4-methyl-2,5-t-hiophene hydrochloride] (PMNT), can catalyze the oxidation reaction of TMB in the presence of H2O2 to produce a blue color solution. The catalytic activity of PMNT on the TMB-H2O2 reaction was closely relevant to the conformation of PMNT. The absorbance of TMB-H2O2 was distinctly increased in the presence of TBA. With the addition of thrombin, TBA interacted with thrombin to form a G-quadruplex structure. The conformational change weakened the catalytic activity of PMNT and resulted in a decrease in the absorbance. The colorimetric sensor could detect thrombin down to 4pM with high selectivity against other interfering proteins. This work is not only of importance for a better understanding of the unique properties of cationic polythiophenes derivative but also have great potential for medical diagnostics and therapy for human health.

关键词