An improved amperometric creatinine biosensor based on nanoparticles of creatininase, creatinase and sarcosine oxidase.

Analytical biochemistry Kumar P, Jaiwal R, Pundir CS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

An improved amperometric biosensor for detection of creatinine was developed based on immobilization of nanoparticles (NPs) of creatininase (CA), creatinase (CI), and sarcosine oxidase (SOx) onto glassy carbon (GC) electrode. Transmission electron microscopy (TEM) and fourier transform infrared spectroscopy (FTIR) were employed for characterization of enzyme nanoparticles (ENPs). The GC electrode was characterized by scanning electron microscopy (SEM), cyclic voltammetry (CV) and electrochemical impedance spectra (EIS) at different stages of its amendment. The biosensor showed optimum response within 2s at pH 6.0 in 0.1 M sodium phosphate buffer and 25 °C, when operated at 1.0 V against Ag/AgCl. Biosensor exhibited wider linear range from 0.01 μM to 12 μM with a limit of detection (LOD) of 0.01 μM. The analytical recoveries of added creatinine in sera were 97.97 ± 0.1% for 0.1 mM and 98.76 ± 0.2% for 0.15 mM, within and between batch coefficients of variation (CV) were 2.06% and 3.09% respectively. A good correlation (R2 = 0.99) was observed between sera creatinine values obtained by standard enzymic colorimetric method and the present biosensor. This biosensor measured creatinine level in sera of apparently healthy subjects and persons suffering from renal and muscular dysfunction. The ENPs electrode lost 10% of its initial activity within 240 days of its regular uses, when stored at 4 °C.