2017

Development and validation of an impedance biosensor for point-of-care detection of vascular cell adhesion molecule-1 toward lupus diagnostics.

Future science OA Selvam AP, Wangzhou A, Jacobs M, Wu T, Mohan C, Prasad S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

AIM: Systemic lupus erythematosus is an autoimmune disease that requires chronic monitoring. In this study, we demonstrate a proof-of-concept study of a highly attractive noninvasive strategy for monitoring systemic lupus erythematosus through biomarker quantification. RESULTS: This sensor technology requires 50 μl of urine to detect and quantify vascular cell adhesion molecule-1 in 15 min. The sensor used nonfaradaic detection to demonstrate performance with and without detection antibody. Binding of immunoassay and target biomarkers were quantified with an impedance electrical immunoassay and correlated with an equivalent circuit. CONCLUSION: The novel sensor technology demonstrates detection in the range of 8 fg/ml to 800 pg/ml and comparative analysis with ELISA platforms was performed for 12 patient urine samples.