2018

High sensitivity cardiac troponin I detection in physiological environment using AlGaN/GaN High Electron Mobility Transistor (HEMT) Biosensors.

Biosensors & bioelectronics Sarangadharan I, Regmi A, Chen YW, Hsu CP, Chen PC, Chang WH, Lee GY, Chyi JI, Shiesh SC, Lee GB, Wang YL
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this study, we report the development of a high sensitivity assay for the detection of cardiac troponin I using electrical double layer gated high field AlGaN/GaN HEMT biosensor. The unique gating mechanism overcomes the drawback of charge screening seen in traditional FET based biosensors, allowing detection of target proteins in physiological solutions without sample processing steps. Troponin I specific antibody and aptamer are used as receptors. The tests carried out using purified protein solution and clinical serum samples depict high sensitivity, specificity and wide dynamic range (0.006-148ng/mL). No additional wash or sample pre-treatment steps are required, which greatly simplifies the biosensor system. The miniaturized HEMT chip is packaged in a polymer substrate and easily integrated with a portable measurement unit, to carry out quantitative troponin I detection in serum samples with < 2µl sample volume in 5min. The integrated prototype biosensor unit demonstrates the potential of the method as a rapid, inexpensive, high sensitivity CVD biomarker assay. The highly simplified protocols and enhanced sensor performance make our biosensor an ideal choice for point of care diagnostics and personal healthcare systems.

关键词