2018

Fully nozzle-jet printed non-enzymatic electrode for biosensing application.

Journal of colloid and interface science Bhat KS, Ahmad R, Yoo JY, Hahn YB
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The progress in developing the electrochemical sensors for biomolecule detection requires a facile device fabrication method. Herein, we report printing of silver (Ag) precursor and copper oxide nanoparticles (CuO NPs) inks by nozzle-jet technique to fabricate non-enzymatic glucose biosensor on flexible polyethylene terephthalate (PET) substrate. The fully printed CuO NPs/Ag/PET electrodes were characterized using electrochemical techniques for non-enzymatic biosensing of glucose. The fully printed non-enzymatic biosensor exhibited a high sensitivity (1424.2 μA mM-1 cm-2), linear range from 0.1 to 15 mM, low detection limit (0.3 μM; S/N = 3) and fast response time of ∼2 s under a working potential of +0.6 V. Additionally, printed electrodes demonstrated an excellent long term stability, high reproducibility, good selectivity and high accuracy during glucose concentration measurements in human blood and serum samples. These results indicate that the electrode fabrication using nozzle-jet printing can be regarded as a potential technique for the future development of high performance and low cost bio/chemical sensor devices.

关键词