2018

Detection of early stage prostate cancer by using a simple carbon nanotube@paper biosensor.

Biosensors & bioelectronics Ji S, Lee M, Kim D
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This study is an investigation for an inexpensive, simple and sensitive biosensor to detect prostate cancer using bioactivated-multi wall carbon nanotubes (MWCNTs, diameter of 20nm, length of 5µm) and a micro-pore filter paper (pore size of 0.45µm). For the immunoassay of prostate specific antigen (PSA), which is a biomarker of prostate cancer, MWCNTs were activated with PSA antibody (monoclonal antibody of the prostate specific antigen) by using N-(3-dimethylaminopropyl)-N'-ethylcarbodiimide hydrochloride (EDC) and N-hydroxysulfosuccinimide sodium salt (NHSS). The activated MWCNTs were deposited on the micro-pore filter paper to use as a biosensor. The prepared biosensor can assay from 0 to 500ng/mL of PSA level within 2h with the detection limit of 1.18ng/mL by the measurement of resistance change. The resistance change was caused by site selective interaction between PSA and PSA-antigen with an inexpensive bench top digital multimeter (5 1/2 digits). The detection range and sensitivity of the prepared sensor are good enough to diagnose the early stage of prostate cancer (> 4ng/mL of PSA). This paper-based biosensor is about 20 times cheaper (fabricated biosensor price: 2.4 $) and over 10 times faster than enzyme-linked immunosorbent assay (ELISA), which is a general method for the detection of a specific protein in the modernized hospitals. Furthermore, the maximum detection limit is about 50 times higher than ELISA.

关键词