2018

SPR imaging biosensor for the quantitation of fibronectin concentration in blood samples.

Journal of pharmaceutical and biomedical analysis Sankiewicz A, Romanowicz L, Pyc M, Hermanowicz A, Gorodkiewicz E
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The purpose of this study was presentation of a new biosensor capable of determination of fibronectin. This biosensor was based on the specific interaction of anti-fibronectin antibody produced in rabbit with fibronectin. The surface plasmon resonance imaging (SPRI) technique was used as a detecting method. Optimization and characterization properties of the biosensor were studied. The determination of fibronectin concentration in natural samples was done. The results were compared with a reference method (Enzyme-Linked Immunosorbent Assay-ELISA). The analytically useful dynamic response range of biosensor is between 5 and 400ngmL-1. The detection limit is 1.5ngmL-1 and limit quantification is 5ngmL-1. The proposed SPRI biosensor showed good selectivity for potential interferences. It was applied to determine fibronectin concentrations in plasma of healthy donors and of patients after thermal injury. Good correlations between results obtained using the SPRI biosensor and ELISA test (correlation coefficients for healthy donors 0.996, for patients 0.984) were obtained. The average fibronectin concentration of healthy donors was 140.5±24.6μgmL -1 and the average fibronectin concentration of patients was 601.5±72.1μgmL -1, which was in agreement with results obtained by other investigators. The obtained results indicate that the developed biosensor may be a candidate for monitoring fibronectin concentration in blood samples.

关键词