Real-time CRP detection from whole blood using micropost-embedded microfluidic chip incorporated with label-free biosensor.

The Analyst Tsai MZ, Hsiung CT, Chen Y, Huang CS, Hsu HY, Hsieh PY
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We demonstrate the detection of C-creative protein (CRP) from whole blood samples without sample pretreatment by using a lab-on-a-chip system consisting of a microfluidic chip and a label-free biosensor. The microfluidic chip includes an array of microposts for filtering blood cells and allows only plasma to flow through and reach the guided-mode resonance (GMR) biosensor for real-time monitoring. The developed GMR sensor can achieve a bulk sensitivity of 186 nm RIU-1, which supports a limit of detection of 3.2 ng mL-1 for recombinant CRP spiked in human serum. The results are comparable with those obtained using enzyme-linked immunosorbent assay. In addition, we demonstrate the efficacy of filtration of blood cells using microposts and simultaneous measurement of CRP concentration using a GMR sensor by using whole blood and plasma samples.

关键词