Fabrication of Electrochemical Biosensor with ZnO-PVA Nanocomposite Interface for the Detection of Hydrogen Peroxide.

Journal of nanoscience and nanotechnology Sekar NK, Gumpu MB, Ramachandra BL, Nesakumar N, Sankar P, Babu KJ, Krishnan UM, Rayappan JBB
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Hydrogen peroxide (H2O2) is considered to be highly toxic and its increased concentration in human body may lead to diseases like alzheimer's, parkinson's, cardiovascular, tumor and cancer. Hence, there is an increasing demand for the detection of H2O2 in human blood serum. In this context, an electrochemical sensor was developed using zinc oxide-polyvinyl alcohol (ZnO-PVA) nanocomposite as a nano-interface. The fabricated Au/ZnO-PVA/CAT/Chitosan bio-electrode exhibited a well-defined redox peak with anodic and cathodic peak potential of -0.408 V and 0.259 V for Fe(III):Fe(II) and H2O2:1/2 O2 redox couples respectively. The developed biosensor exhibited a linear range of 1 μM-17 μM with a sensitivity of 210.49 μA μM-1 cm-2, response time of less than 1 s, limit of detection of 9.13 nM and a limit of quantification of 30.13 nM. The developed bio-electrode showed a Michaelis-Menten constant (KM) of 0.39 μM and dry stability of 93% up to 20 days. The obtained biosensor was successfully utilized to determine the H2O2 concentration in human blood serum sample.