Problems analysis and new fabrication strategies of mediated electrochemical biosensors for wastewater toxicity assessment.

Biosensors & bioelectronics Yang Y, Fang D, Liu Y, Liu R, Wang X, Yu Y, Zhi J
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Conventional mediated electrochemical biosensors for toxicity assessment were almost based on 'one-pot' principle, i.e., mediators and the under-test chemicals were mixed together in the same vessel. In this process, the electron mediator is assumed to be merely an electron acceptor and cannot react with under-test toxicants. Actually,some under-test pollutants (such as metal ions) could react with the electron mediators, thus affecting the detection accuracy and sensitivity of the sensors. It was also found that at least two other interference factors have been ignored in present'one-pot' mediated electrochemical biosensor systems, i.e., (1) the electrochemical sensitivity of mediators to pH; and (2) the potential reactions between under-test chemicals and buffers and the consequent pH changes. In this study, the three ignored interference factors have been investigated systematically and demonstrated by significance tests. Moreover, a solving strategy, an isolation method, is proposed for fabrication of novel mediated electrochemical biosensor to avoid the interference factors existing at present mediated electrochemical biosensor. According to the testing results obtained from the isolation method, IC50 values of Cu2+, Cd2+, Zn2+, Fe3+, Ni2+ and Cr3+ were 21.3 mg/L, 3.7 mg/L, 26.7 mg/L, 4.4 mg/L and 10.7 mg/L, respectively. The detection results of four real water samples also suggested this method could be applied for the practical and complex samples.

关键词