2018

A label-free fluorescent biosensor for the detection of protein kinase activity based on gold nanoclusters/graphene oxide hybrid materials.

Analytica chimica acta Liu Q, Li N, Wang M, Wang L, Su X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Protein kinase (PKA) can regulate many cellular biological processes by phosphorylation substrate peptide or protein. A new fluorescent biosensing method for the detection of PKA activity was developed by using 11-mercaptoundecanoic acid-capped gold nanoclusters (MUA-Au NCs) and graphene oxide (GO) with low background noise. In this strategy, the special designed peptide could be anchored on the surface of MUA-Au NCs by the Au-S bond and also adsorbed on the surface of GO owing to the electrostatic interaction. As a result, the fluorescence of MUA-Au NCs was quenched leading to low background fluorescence due to the forster resonance energy transfer (FRET) between MUA-Au NCs and GO via peptide as a bridge. However, when the substrate peptide was phosphorylated by PKA, the FRET between GO and MUA-Au NCs was disrupted because of the weakened interaction between the phosphorylated peptide and the GO, resulting in recovery of the fluorescence intensity. The developed label-free fluorescence "turn-off-on" method can detect protein kinase activity in the range of 0.6-2.0 U mL-1 with a detection limit of 0.17 U mL-1 (3σ). The feasibility of this present method for kinase inhibitor screening was also studied by assessment of H-89 kinase inhibition with an IC50 value of 0.049 μmol L-1.

关键词