Miniaturized Reverse Electrodialysis-Powered Biosensor Using Electrochemiluminescence on Bipolar Electrode.

Analytical chemistry Baek S, Kwon SR, Yeon SY, Yoon SH, Kang CM, Han SH, Lee D, Chung TD
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We suggest an electrochemiluminescence (ECL)-sensing platform driven by ecofriendly, disposable, and miniaturized reverse electrodialysis (RED) patches as an electric power source. The flexible RED patches composed of ion-exchange membranes (IEMs) can produce voltage required for ECL sensing by simply choosing the appropriate number of IEMs and the ratio of salt concentrations. We integrate the RED patch with a bipolar electrode on the microfluidic chip to demonstrate the proof-of-concept, i.e., glucose detection in the range of 0.5-10 mM by observing ECL emissions with naked eyes. The miniaturized RED-powered biosensing system is widely applicable for electrochemical-sensing platforms. This is expected to be a solution for practical availability of battery-free electrochemical sensors for disease diagnosis in developing countries.