2018

Integrated Biosensor for Rapid and Point-of-Care Sepsis Diagnosis.

ACS nano Min J, Nothing M, Coble B, Zheng H, Park J, Im H, Weber GF, Castro CM, Swirski FK, Weissleder R, Lee H
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Sepsis is an often fatal condition that arises when the immune response to an infection causes widespread systemic organ injury. A critical unmet need in combating sepsis is the lack of accurate early biomarkers that produce actionable results in busy clinical settings. Here, we report the development of a point-of-care platform for rapid sepsis detection. Termed IBS (integrated biosensor for sepsis), our approach leverages (i) the pathophysiological role of cytokine interleukin-3 (IL-3) in early sepsis and (ii) a hybrid magneto-electrochemical sensor for IL-3 detection. The developed platform produces test results within 1 h from native blood samples and detects IL-3 at a sensitivity of <10 pg/mL; this performance is >5-times faster and >10-times more sensitive than conventional enzyme-linked immunoadsorbent assays, the current gold standard. Using clinical samples, we show that elevated plasma IL-3 levels are associated with high organ failure rate and thus greater risk of mortality, confirming the potential of IL-3 as a sepsis diagnostic biomarker. With further system development ( e. g., full automation, data security measures) and rigorous validation studies, the compact and fast IBS could be a practical clinical tool for timely diagnosis and proactive treatment of sepsis.

关键词