Sensitive electrogenerated chemiluminescence biosensors for protein kinase activity analysis based on bimetallic catalysis signal amplification and recognition of Au and Pt loaded metal-organic frameworks nanocomposites.

Biosensors & bioelectronics Yan Z, Wang F, Deng P, Wang Y, Cai K, Chen Y, Wang Z, Liu Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work, a novel and sensitive electrogenerated chemiluminescence (ECL) biosensor for protein kinase A (PKA) activity analysis and relevant inhibitor screening was proposed based on bimetallic catalysis signal amplification and recognition of Au and Pt nanoparticles loaded metal-organic frameworks (Au&Pt@UiO-66) nanocomposite. After being phosphorylated by PKA in the presence of ATP, Au&Pt@UiO-66 probes were specifically chelated to the modified electrode by forming Zr-O-P bonds between the surface defects of UiO-66 and the phosphorylated kemptide. Due to the high synergistic catalysis of Au&Pt@UiO-66 nanocomposites to the luminol-H2O2 reaction, the ECL signal of luminol was greatly enhanced. Moreover, UiO-66 afford numerous Zr defect sites for high efficient phosphate group recognition, and can also prevent the nanoparticles from aggregating during catalytic reactions. Thus, the excellent performance of the ECL biosensor with high sensitivity and superior stability was obtained. Under the optimized conditions, the detection limit for PKA activity was 0.009 UmL-1 (S/N = 3). Meanwhile, the ECL biosensor was successfully applied in inhibitor screening and cell lysates PKA activity analysis, showing great promise in kinase related research.