2018

Development of Cu nanoflowers modified the flexible needle-type microelectrode and its application in continuous monitoring glucose in vivo.

Biosensors & bioelectronics Fang Y, Wang S, Liu Y, Xu Z, Zhang K, Guo Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A minimally invasive glucose microbiosensor based the flexibly integrated electrode for continuous monitoring glucose in vivo has been developed in this study. This was achieved by coating needle-type microelectrode with Cu nanoflowers, nafion, glucose oxidase (GOD) and polyurethane (PU) membranes, successfully prepared with layer-by-layer deposition. The Cu nanomaterials provided a large specific surface area and electrocatalytic activity for glucose detection. The PU layers as mass-transport limiting membranes significantly enhanced the linearity and stability of sensors. The resulting biosensor exhibited a wide linear range of 0-20 mM, with a good sensitivity of 42.38 nA mM-1 (correlation coefficient r2 was 0.99) and a fast response time of less than 15 s. In vivo implantable experiments using anesthetized rats showed excellent real-time response to the variation of blood glucose concentration. And the variation tendency of sensor output was consistent with that using the glucose meter. Overall, the results supported the suitability of this microsensor for measuring rapid changes of glucose in vivo. This work offers a promising approach in implantable device applications related to diabetes management as well as other medical diagnosis.

关键词