2017

Electrochemical immunosensor for the breast cancer marker CA 15-3 based on the catalytic activity of a CuS/reduced graphene oxide nanocomposite towards the electrooxidation of catechol.

Mikrochimica acta Amani J, Khoshroo A, Rahimi-Nasrabadi M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The authors report on an electrochemical immunosensor for the tumor marker carbohydrate antigen 15-3 (CA15-3). It is based on the use of a composite consisting of reduced graphene oxide (RGO) and copper sulfide (CuS) that was placed on a screen-printed graphite electrode. The electrode shows excellent activity towards the oxidation of catechol acting as an electrochemical probe, best at a working potential of 0.16 V. The electrode was modified with antibody against CA15-3. Once the analyte (CA15-3) binds to the surface of the electrode, the response to catechol is reduced. The assay has a linear response in the 1.0-150 U mL-1 CA15-3 concentration range, with a 0.3 U mL-1 lower detection limit and a sensitivity of 1.88 μA μM-1 cm-2. The immunosensor also shows good reproducibility (2.7%), stability (95% of the initial values after storing for four weeks). The method was successfully applied to the determination of CA15-3 in serum samples, and results were found to compare well to those obtained by an ELISA. Conceivably, this nanocomposite based detection scheme has a wider scope and may be applied to numerous other immunoassays. Graphical abstract A label-free electrochemical immunosensor based on copper sulfides/graphene nanocomposites was developed for enzyme-free determination of CA15-3 biomarker. This immunosensor can be utilized as a tool to detect of CA15-3 in real samples.