2018

A Disposable Tear Glucose Biosensor-Part 5: Improvements in Reagents and Tear Sampling Component.

Journal of diabetes science and technology Lin CE, Ito Y, Deng A, Johns J, Matloff D, Cook CB, Sode K, La Belle JT
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A tear glucose (TG) sensor with an integrated tear sampler can provide a noninvasive method for calibrating the continuous TG contact lens and monitoring glucose. Expanding from previous work, an improved TG sensor that implements dried reagents, genetically modified glucose dehydrogenase (GDH), and a tear sampler was developed and compared against the TG sensor prepared with commercial GDH. It was found that neither sensor was affected by the tear interferents: ascorbic acid, acetaminophen, and uric acid. The sensor prepared with commercial GDH generated higher current. This suggests that using enzymes with lower Km may be advantageous when operating in low glucose environments like tears. The improved TG sensor also demonstrated the potential of integrating Schirmer's test strip as a tear sampler for self-monitoring of TG.