2018

An interdigitated electrode biosensor platform for rapid HLA-B*15:02 genotyping for prevention of drug hypersensitivity.

Biosensors & bioelectronics Soraya GV, Chan J, Nguyen TC, Huynh DH, Abeyrathne CD, Chana G, Todaro M, Skafidas E, Kwan P
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Prevention of life threatening hypersensitivity reactions to carbamazepine is possible through pre-treatment screening of the associated HLA-B*15:02 risk allele. However, clinical implementation of screening is hindered by the high cost and slow turnaround of conventional HLA typing methods. We have developed an interdigitated electrode (IDE) biosensor platform utilizing loop mediated isothermal amplification (LAMP) that can rapidly detect the HLA-B*15:02 allele. DNA amplification is followed by solid-phase hybridization of LAMP amplicons to a DNA probe immobilized on the IDE sensor surface, resulting in a change in sensor impedance. The testing platform does not require DNA extraction or post-amplification staining, achieving sample-to-answer in 1 h and 20 min. The platform was tested on 27 whole blood samples (14 HLA-B*15:02 positive and 13 negative) with sensitivity of 92.9% and specificity of 84.6% when applying a cutoff of impedance change. Based on these characters the LAMP-IDE platform has potential to be further developed into point-of-care use to help overcome barriers in HLA-B*15:02 screening.

关键词