2018

Amperometric determination of As(III) and Cd(II) using a platinum electrode modified with acetylcholinesterase, ruthenium(II)-tris(bipyridine) and graphene oxide.

Mikrochimica acta Gumpu MB, Veerapandian M, Krishnan UM, Rayappan JBB
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The authors describe an amperometric biosensor for the determination As(III) and Cd(II) based on the inhibition of the enzyme acetylcholineesterase (AChE). A platinum electrode was modified with ruthenium(II)-tris(bipyridyl), graphene oxide and AChE and then showed redox peaks at 0.06 and 0.2 V vs Ag/AgCl in the presence of acetylthiocholine chloride (ATChCl). Amperometry unveiled a steady-state turnover rate with the release of thiocholine. In the presence of arsenic(III) and cadmium(II), AChE showed an inhibitive response at 0.214 and 0.233 V vs Ag/AgCl, respectively. The electrode exhibits a detection limit and linear range of 0.03 μM and 0.05-0.8 μM for As(III) and 0.07 μM and 0.02-0.7 μM for Cd(II), respectively. Type of inhibition and inhibition constants induced by As(III) and Cd(II) on the catalytic sites of AChE were determined from Dixon and Lineweaver-Burk plots. The modified electrode was applied to the determination of As3+ and Cd2+ in river, tap and waste water, and the results proved that the method is sensitive and can be an alternative to chromatographic and spectroscopic techniques. Graphical abstract Schematic presentation of Pt/Ru(II)-tris(bipy)-GO/AChE electrode in absence and presence of metal ions (As3+/Cd2+).

关键词