2018

Graphene oxide-assisted Au nanoparticle strip biosensor based on GR-5 DNAzyme for rapid lead ion detection.

Colloids and surfaces. B, Biointerfaces Wang HB, Ma LH, Fang BY, Zhao YD, Hu XB
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This study has reported that a GR-5 DNAzyme based lead ion strip biosensor could exhibit an enhanced specificity with the assistance of graphene oxide (GO). This enhancement results from the specific π-stacking interaction between the ribose rings of the nucleobases and the carbon hexagons in GO which can reduce the false positive interference by removing unhybridized ssDNA during the annealing of GR-5 DNAzyme. Meanwhile, conjugate pad was sprayed with two kinds of AuNP-DNA probes, and nitrocellulose membrane test zone and control zone were immobilized with two kinds of biotin-DNA probes, respectively. The limit of detection of this strip biosensor was estimated to be about 0.05 nM (S/N = 3) and 1 nM (with naked eyes) with a linear range from 0.01 to 100 μM. Furthermore, the strip biosensor exhibited excellent selectivity toward Pb2+ in the presence of other divalent metal ions. For real soil samples, the obtained recoveries were in the range from 91.5% to 113.1%.

关键词