2018

Development of an Amperometric Glucose Biosensor Based on the Immobilization of Glucose Oxidase on the Se-MCM-41 Mesoporous Composite.

Journal of analytical methods in chemistry Yusan S, Rahman MM, Mohamad N, Arrif TM, Latif AZA, M A MA, Wan Nik WSB
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A new bioenzymatic glucose biosensor for selective and sensitive detection of glucose was developed by the immobilization of glucose oxidase (GOD) onto selenium nanoparticle-mesoporous silica composite (MCM-41) matrix and then prepared as a carbon paste electrode (CPE). Cyclic voltammetry was employed to probe the catalytic behavior of the biosensor. A linear calibration plot is obtained over a wide concentration range of glucose from 1 × 10-5 to 2 × 10-3 M. Under optimal conditions, the biosensor exhibits high sensitivity (0.34 µA·mM-1), low detection limit (1 × 10-4 M), high affinity to glucose (Km = 0.02 mM), and also good reproducibility (R.S.D. 2.8%, n=10) and a stability of about ten days when stored dry at +4°C. Besides, the effects of pH value, scan rate, mediator effects on the glucose current, and electroactive interference of the biosensor were also discussed. As a result, the biosensor exhibited an excellent electrocatalytic response to glucose as well as unique stability and reproducibility.

关键词