Modelling of magnetoimpedance response of thin film sensitive element in the presence of ferrogel: Next step toward development of biosensor for in-tissue embedded magnetic nanoparticles detection.

Biosensors & bioelectronics Buznikov NA, Safronov AP, Orue I, Golubeva EV, Lepalovskij VN, Svalov AV, Chlenova AA, Kurlyandskaya GV
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In-tissue embedded magnetic nanoparticle (MNPs) detection is one of the most interesting cases for cancer research. In order to understand the origin, the limits and the way of improvement of magnetic biosensor sensitivity for the detection of 3D mezoscopic distributions of MNPs, we have developed a magnetoimpedance biosensor prototype with a [Cu (3 nm)/FeNi(100 nm)]5/Cu(500 nm)/[FeNi(100 nm)/Cu(3 nm)]5 rectangular sensitive element. Magnetoimpedance (MI) responses were measured with and without polyacrylamide ferrogel layer mimicking natural tissue in order to evaluate stray fields of embedded MNPs of γ-Fe2O3 iron oxide. A model for MI response based on a solution of Maxwell equations with Landau-Lifshitz equation was developed in order to understand the origin of the prototype sensitivity which reached 1.3% of ΔZ/Z per 1% of MNPs concentration by weight. To make this promising technique useful for magnetically labeled tissue detection, a synthesis of composite gels with MNPs agglomerates compactly located inside pure gel and their MI testing are still necessary.