2018

Xurography-based microfluidic algal biosensor and dedicated portable measurement station for online monitoring of urban polluted samples.

Biosensors & bioelectronics Gosset A, Durrieu C, Renaud L, Deman AL, Barbe P, Bayard R, Chateaux JF
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A critical need exists to develop rapid, in situ, and real-time tools to monitor the impact of pollution discharge toxicity on aquatic ecosystems. The present paper deals with the development of a novel, simple-to-use, low-cost, portable, and user-friendly algal biosensor. In this study, a complete and autonomous portable fluorimeter was developed to assess the A-chlorophyll fluorescence of microalgae, inserted by capillarity into low-cost and disposable xurography-based microfluidic chips. Three microalgae populations were used to develop the biosensor: Chlorella vulgaris, Pseudokirchneriella subcapitata, and Chlamydomonas reinhardtii. Biosensor feasibility and sensitivity parameters, such as algal concentration and light intensity, were optimized beforehand to calibrate the biosensor sensitivity with Diuron, a pesticide known to be very toxic for microalgae. Finally, the biosensor was employed in 10 aqueous urban polluted samples (7 urban wet-weather discharges and 3 wastewater) in order to prove its reliability, reproducibility, and performance in the detection of toxic discharges in the field.