2018

Conductometric biosensor for arginine determination in pharmaceutics.

Bioelectrochemistry (Amsterdam, Netherlands) Soldatkina OV, Soldatkin OO, Velychko TP, Prilipko VO, Kuibida MA, Dzyadevych SV
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A new conductometric biosensor based on coimmobilized urease and arginase has been developed for arginine determination in pharmaceutics. First, the main parameters of the selected method of immobilization (concentrations of arginase, urease, and glutaraldehyde, time of incubation) were optimized. An influence of the solution parameters (buffer ionic strength, capacity, pH, Mn2+ concentration) on the biosensor operation was studied, working conditions were optimized. After biosensor optimization, the main analytical characteristics were as follows. The limit of detection - 2.5 μM, the linear range - 2.5-500 μM, the sensitivity to arginine 13.4 ± 2.4 μS/mM, the response time - 20 s. The signals repeatability and operational stability in continuous exploitation were studied over one working day and during one week. Additionally, the selectivity of the developed biosensor towards arginine was essayed relative to other amino acids. The developed biosensor has been used to measure arginine concentrations in some drugs. The results obtained were in high correlation with the characteristics declared by producers.

关键词