组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
We have developed a low-cost optical cavity-based biosensor with a differential detection method for point-of-care medical diagnostics. To experimentally demonstrate its label-free real-time biosensing capability, we performed the detection of biotinylated bovine serum albumin (BSA). Streptavidin is introduced into the optical cavity structure and immobilized on 3-aminopropyltriethoxysilane (APTES) coated surface. After rinsing out unbound streptavidin with DI water, biotinylated BSA without any labeling is introduced. A CMOS camera captures the transmitted light of two different wavelengths passing through the optical cavity sensing area in real-time. Then, the differential values are calculated to enhance the responsivity. We successfully demonstrated the label-free real-time detection of biotinylated BSA, and the measurement results matched well with the simulation results. The limit of detection of the optical cavity-based biosensor for the biotinylated BSA detection with the sensing area of 180 μm × 180 μm is estimated to be 2.82 pM, which could be reduced further for a smaller sensing area with the tradeoff of a longer sensing time.