2018

Use of a Turn Coil and Channel Above a GMR-SV Device to Observe and Measure the Properties of Deoxidized Red Blood Cells Coupled to Magnetic Beads.

Advances in experimental medicine and biology Choi JG, Kim SH, Lee SS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A giant magnetoresistance-spin valve (GMR-SV) having highly sensitive magnetic properties was prepared with a multilayer soft ferromagnetic/non-magnetic/pinned ferromagnetic/anti-ferromagnetic iridium manganese (IrMn) top layer (NiFe/CoFe/Cu/CoFe/IrMn) structure. Micron-sized magnetic beads (MBs) with a diameter of 1 μm immersed in a solution containing 50 mg of Co-Si-OH/ml were attached to red blood cells (RBCs) having the very low magnetization of hemoglobin (HEM) to detect a deoxidized RBC combined with MBs (RBC+MBs). The RBC+MB combinations can be captured on a single-turn μ-coil that maintains a magnetic field large enough to detect the MBs attached to the RBC. When the RBC+MBs passes through a PR μ-channel with a width of 10 μm, its stopping and starting are controlled by the electrical AC input signal applied to the single-turn μ-coil. The RBC+MBs combinations captured above the μ-device of the GMR-SV change the output signal and thus, can be used for detection. This observation implies that this device can be used as a biosensor to analyze the coupling force between the HEM and the MBs for RBCs with deformed membranes passing through a narrow capillary.

关键词