In the field of label-free biosensing, various transducer materials and strategies are under investigation to overcome the Debye-screening limitation of charged biomolecules. We demonstrate an in-line, impedimetric aptasensor with reduced graphene-oxide (rGO) thin films as transducers to detect prostate specific antigens (PSA) in a physiological buffer solution. Unlike classical electrochemical impedance spectroscopy (EIS), this direct, label-free and fully-electronic biosensor approach does not need any redox markers. As specific capture molecules, short anti-PSA aptamers ensured a close binding of the target molecules to the transducer surfaces. Results showed a limit of detection smaller than 33 pM of PSA and a wide detection range from 0.033 to 330 nM fully covering the clinically relevant range of PSA (0.115-0.290 nM). This promising performance can be attributed to the bipolar electronic transport characteristics of the ultra-thin rGO layers similar to pristine graphene. The attachment of target biomolecules to the films changes the resistance of the rGO thin films. Such an in-line EIS configuration with rGO thin films opens promising prospects for biosensing beyond the Debye-screening limitation, which is a major challenge for conventional semiconductor field-effect devices towards clinical applications.