2019

Quantification of cDNA on GMR biosensor array towards point-of-care gene expression analysis.

Biosensors & bioelectronics Ravi N, Rizzi G, Chang SE, Cheung P, Utz PJ, Wang SX
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Gene expression analysis at the point-of-care is important for rapid disease diagnosis, but traditional techniques are limited by multiplexing capabilities, bulky equipment, and cost. We present a gene expression analysis platform using a giant magnetoresistive (GMR) biosensor array, which allows multiplexed transcript detection and quantification through cost-effective magnetic detection. In this work, we have characterized the sensitivity, dynamic range, and quantification accuracy of Polymerase chain reaction (PCR)-amplified complementary DNA (cDNA) on the GMR for the reference gene GAPDH. A synthetic GAPDH single-stranded DNA (ssDNA) standard was used to calibrate the detection, and ssDNA dilutions were qPCR-amplified to obtain a standard curve. We demonstrate that the GMR platform provides a dynamic range of 4 orders of magnitude and a limit of detection of 1 pM and 0.1 pM respectively for 15 and 18-cycle amplified synthetic GAPDH PCR products. The quantitative results of GMR analysis of cell-line RNA were confirmed by qPCR.