2018

Ultrasensitive Biosensor for Detection of Mercury(II) Ions Based on DNA-Cu Nanoclusters and Exonuclease III-assisted Signal Amplification.

Analytical sciences : the international journal of the Japan Society for Analytical Chemistry Zhang H, Guan Y, Li X, Lian L, Wang X, Gao W, Zhu B, Liu X, Lou D
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This paper describes a novel method for label-free mercury(II) ion detection based on exonuclease III-induced target signal recycling amplification using double-stranded DNA templated copper nanoclusters. The synthesized DNA-Cu nanoclusters were used with exonuclease III loop amplification technology for ultra-high sensitivity detection of mercury(II) ions, which were detected by significantly decreased fluorescence intensity. Under the optimal experimental conditions, there was a clear linear relationship between Hg2+ concentration in the range of 0.04 to 8 nM and fluorescence intensity. The detection limit for Hg2+ was 4 pM. In addition, the interference of other metal ions on the mercury(II) ion detection was also studied. To confirm the application of the fluorescent sensor, it was applied to determine the concentrations of mercury(II) ions in tap water, and the results showed that the method can be used to detect mercury(II) ions in water samples successfully.

关键词