2018

Non-invasive Point-of-Care Device To Diagnose Acute Mesenteric Ischemia.

ACS sensors Abdelrasoul GN, MacKay S, Salim SY, Ismond KP, Tamura M, Khalifa C, Mannan E, Lin D, Mandal T, Montgomery RR, Wishart DS, Chen J, Khadaroo RG
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Inadequate blood supply to the intestine can lead to acute mesenteric ischemia (AMI), with a mortality rate ranging from 60% to 90%. This high mortality rate is partially due to late detection and the lack of efficient early diagnostic tests. There is an urgent need for a point-of-care tool for immediate bedside diagnosis. Here we present for the first time a rapid and non-invasive electrochemical biosensor device based on non-faradic impedance spectroscopy to detect intestinal fatty-acid binding protein (I-FABP) as an indication of AMI. The electrochemical biosensors consist of gold interdigitated electrodes that were fabricated using photolithographic techniques on top of silicon dioxide substrates. The electrode surfaces were functionalized with an I-FABP capture antibody (CAnB) to entice the target protein, while gold nanoparticles (GNPs) functionalized with detection antibodies (DAnB-GNPs) were utilized as a novel mechanism to enhance the detection signal. Quantification of the I-FABP concentration in the medium depended on its attachment to CAnB and DAnB-GNPs in a sandwich manner, where the latter boosts the impedance signal through its binding to the I-FABP. This non-invasive non-faradic electric biosensor device demonstrates the potential for bench-to-bedside translation with the goal of decreasing morbidity and mortality from AMI.