2019

Monolithic CMOS sensor platform featuring an array of 9'216 carbon-nanotube-sensor elements and low-noise, wide-bandwidth and wide-dynamic-range readout circuitry.

Sensors and actuators. B, Chemical Dudina A, Seichepine F, Chen Y, Stettler A, Hierlemann A, Frey U
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We present the design and characterization of a monolithic complementary metal-oxide-semiconductor (CMOS) biosensor platform comprising of a switch-matrix-based array of 9'216 carbon nanotube field-effect transistors (CNTFETs) and associated readout circuitry. The switch-matrix allows for flexible selection and simultaneous routing of 96 sensor elements to the corresponding readout channels. A low-noise, wide-bandwidth, wide-dynamic-range transimpedance continuous-time amplifier architecture has been implemented to facilitate resistance measurements in the range between 50 kΩ and 1 GΩ at a bandwidth of up to 1 MHz. The achieved accuracy of the resistance measurements over the whole range is 4%. The system has been successfully fabricated and tested and shows a noise performance equal to 2.14 pArms at a bandwidth of 1 kHz and 0.84 nArms at a bandwidth of 1 MHz. A batch integration of the CNTFETs has been achieved by using a dielectrophoresis (DEP)-based manipulation technique. The current-voltage curves of CNTFETs have been acquired, and the sensing capabilities of the system have been demonstrated by recording resistance changes of CNTFETs upon exposure to solutions with different pH values and different concentrations of NaCl. The smallest resolvable concentrations for the respective analytes were estimated to amount to 0.025 pH-units and 4 mM NaCl.

关键词