2018

Quantification of DNA by a Thermal-Durable Biosensor Modified with Conductive Poly(3,4-ethylenedioxythiophene).

Sensors (Basel, Switzerland) Gu Y, Tseng PY, Bi X, Yang JHC
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The general clinical procedure for viral DNA detection or gene mutation diagnosis following polymerase chain reaction (PCR) often involves gel electrophoresis and DNA sequencing, which is usually time-consuming. In this study, we have proposed a facile strategy to construct a DNA biosensor, in which the platinum electrode was modified with a dual-film of electrochemically synthesized poly(3,4-ethylenedioxythiophene) (PEDOT) resulting in immobilized gold nanoparticles, with the gold nanoparticles easily immobilized in a uniform distribution. The DNA probe labeled with a SH group was then assembled to the fabricated electrode and employed to capture the target DNA based on the complementary sequence. The hybridization efficiency was evaluated with differential pulse voltammetry (DPV) in the presence of daunorubicin hydrochloride. Our results demonstrated that the peak current in DPV exhibited a linear correlation the concentration of target DNA that was complementary to the probe DNA. Moreover, the electrode could be reused by heating denaturation and re-hybridization, which only brought slight signal decay. In addition, the addition of the oxidized form of nicotinamide adenine dinucleotide (NAD⁺) could dramatically enhance the sensitivity by more than 5.45-fold, and the limit-of-detection reached about 100 pM.

关键词