2018

Simple, Fast And Highly Sensitive Detection Of Gram-Negative Bacteria By A Novel Electrical Biosensor.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Wu J, Cheng C, Yuan Q, Oueslati R, Zhang J, Chen J, Almeida R
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This work presents a rapid, low-cost, highly sensitive and specific capacitive sensor for detection of Gram negative bacteria in a field setting. Recognition of Gramnegative bacteria is based on specific detection of lipopolysaccharides (LPS) by LPS-specific aptamer probe immobilized on electrode sensors. An inhomogeneous AC electric field is applied on sensor electrodes and induces positive dielectrophoresis that attracts LPS particles to the sensor electrodes for rapid detection. The same AC signal is also used to detect the binding reactions occurred on the sensor surface. The AC signal was optimized, and the binding between LPS and the specific aptamer was demonstrated. The detection limit reaches as low as 4.9 fg/mL for free LPS molecules and 53 #/mL of bacteria within a 30s' response time, meeting the needs of onsite bacteria detection.

关键词