Fully Disposable Wireless Patch Sensor for Continuous Remote Patient Monitoring.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Selvaraj N, Nallathambi G, Moghadam R, Aga A
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Continuous remote monitoring with convenient wireless sensors is attractive for early detection of patient deterioration, preventing adverse events and leading to better patient care. This article presents an innovative sensor design of VitalPatch, a fully disposable wireless biosensor, for remote continuous monitoring, and details the performance assessments from bench testing and laboratory validation in 57 subjects. The bench testing results reveal that VitalPatch's QRS detection had a positive predictive value of $> 99$% from testing with ECG databases. The accuracies of HR, BR and skin temp (in mean absolute error, MAE) from bench testing were $< 5$ bpm, $< 1$ brpm, $< 1 ^{ \circ}C$ respectively. The laboratory testing in 57 subjects revealed the accuracy of HR and BR to be $2.2 \pm 1.5$ bpm and $1.7 \pm 0.7$ brpm respectively for stationary periods. The absolute percent error in detecting steps was $4.7 \pm 4.6$%, and the accuracy in detecting posture was $96.4 \pm 3.1$%. Meanwhile, the specificity and sensitivity of fall detection $( \mathrm {n}=20)$ was found to be 100% and 93.8%, respectively. In conclusion, VitalPatch biosensor demonstrated clinically acceptable accuracies for its vital signs and actigraphy metrics applicable for continuous unobtrusive patient monitoring.