2019

Development of a multiplexed giant magnetoresistive biosensor array prototype to quantify ovarian cancer biomarkers.

Biosensors & bioelectronics Klein T, Wang W, Yu L, Wu K, Boylan KLM, Vogel RI, Skubitz APN, Wang JP
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work, we developed benchtop and handheld Giant Magnetoresistive (GMR) biosensing systems that serve as platforms for detecting a wide variety of protein biomarkers for human diseases. System development included spintronic and nanomagnetic materials, biomolecular chemistry, electronic circuitry, analog and digital signal processing, firmware programming, user interface programming on both PC and Android smartphone, communications over both USB and Bluetooth, and mechanical integration. In this work, we demonstrated the benchtop GMR biosensing system in the context of ovarian cancer assay development. The prototype system delivered the required performance in terms of high-sensitivity multiplex assays in a portable format with enough flexibility to serve as a platform for ovarian cancer and many other diseases. We achieved multiplex detection of cancer antigen 125 (CA125 II), human epididymis protein 4 (HE4), and interleukin 6 (IL6), with limits of detection (LOD) as low as 3.7 U/mL, 7.4 pg/mL, and 7.4 pg/mL, respectively.

关键词