This study compares the biosensing performance of a microbial fuel cell (MFC) and a microbial electrolysis cell (MEC). Initial tests provided a qualitative comparison of MFC and MEC currents after the anode compartment liquid (anolyte) was spiked with acetate, or sulphates of NH4+, Na+, Mg2+, Fe2+, or a fertilizer solution. Current measurements showed that the MFC sensor had a faster response time, higher sensitivity, and faster recovery time after the spike. Following the spike tests, the MFC and MEC were operated in a continuous flow mode at several influent concentrations of acetate, and sulphates of NH4+, Na+, and Fe2+. The continuous flow tests confirmed the better performance of the MFC sensor, which was selected for further experiments. Two MFC sensors were used for real-time (on-line) COD measurements of brewery wastewater. Regression analysis showed a strong correlation between the MFC power output and COD concentrations in the anode compartment with a coefficient of determination (R2) of 0.97. Overall, results of this study suggest that an MFC-based sensor can be successfully used as a simple and cost-efficient real-time monitoring tool.