2019

Dual-Wavelength Ratiometric Electrochemiluminescence Immunosensor for Cardiac Troponin I Detection.

Analytical chemistry Ye J, Zhu L, Yan M, Zhu Q, Lu Q, Huang J, Cui H, Yang X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Ratiometric electrochemiluminescence (ECL) has attracted special focus in the biological analysis field, because it could eliminate the environmental interference and allow for precise measurement. Herein, a dual-wavelength ratiometric ECL biosensor was designed for the detection of cardiac troponin I (cTnI), where (4,4'-dicarboxylic acid-2,2'-bipyridyl) ruthenium(II) (Ru(dcbpy)32+) and Au nanoparticle-loaded graphene oxide/polyethylenimine (GPRu-Au) nanomaterial acts as an acceptor, and Au nanoparticle-modified graphitic phase carbon nitride nanosheet composite (Au-CNN) acts as donor. Au-CNN shows a high and steady ECL signal centered at 455 nm, which is well-matched with the adsorption of GPRu-Au; thereby, a highly efficient electrochemiluminescent resonance energy transfer (ECL-RET) sensing platform is designed. AuNPs facilitate the immobilization of antibody on the nanomaterials through a Au-N bond. The high surface area of graphene oxide/polyethylenimine allows a large number of Ru(dcbpy)32+ to be loaded, immensely amplifying the ECL signal. This sensing platform exhibits outstanding analytical performance toward cTnI with a detection limit of 3.94 fg/mL (S/N = 3). The high reliability, selectivity, and sensitivity of this ratiometric ECL biosensor provides a versatile sensing platform for the bioanalysis.

关键词