Use of zein microspheres to anchor carbon black and hemoglobin in electrochemical biosensors to detect hydrogen peroxide in cosmetic products, food and biological fluids.

Talanta Dos Santos Pereira T, Mauruto de Oliveira GC, Santos FA, Raymundo-Pereira PA, Oliveira ON, Janegitz BC
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Hemoglobin-containing electrochemical biosensors are useful for detecting hydrogen peroxide through oxidation of the iron ion, but high efficiency can only be reached with appropriate immobilization strategies for hemoglobin. In this work, we combined zein from corn seed with carbon black to immobilize hemoglobin, as proof of concept, and form an electroactive film that could determine hydrogen peroxide within the concentration range from 4.9 × 10-6 to 3.9 × 10-4 mo L-1, and limit of detection of 4.0 × 10-6 mol L-1, using differential pulse voltammetry. The biosensor could also detect hydrogen peroxide in commercial samples of oxygenated water, synthetic serum (physiological and glycoside) and milk. The high performance is ascribed to the large surface area and conductive nature of the porous film that had carbon black and hemoglobin anchored on zein microspheres, according to scanning and transmission electron microscopies. It is significant that a protein from renewable sources (zein) combined with a low-cost carbon material (carbon black) serves as matrix for immobilization of biomolecules.